Teledyne 7600 - Infrared gas analyzer - operating manual User Manual

Page 29

Advertising
background image

Infrared Gas Analyzer

Installation

Teledyne

Analytical

Instruments

29

• The dew point of the sample gas must be lower th

an

the

ambient temperature to avoid condensation in the analyzer. If
vapor is contained in the sampling gas, the dew point should
be lowered to 0°C by using a dehumidifier.

• If SO

2

mist is contained in the sample gas, use a mist filter or

cooler to remove the SO

2

mist. Other mists should be

removed by using an appropriate mist filter or cooler.

• Corrosive gases such as Cl

2

, F

2

an

d

HCl, if contained in the

sample gas in considerable amounts, will sho

rt

en the life of

the instrument.

• Sample gas temperature should be within 0 to 50°C. Hot gas

should not be fed directly into the instrument.

2.5.4 Flowrate

Use a flowrate of 0.5L/min ±0.2L/min. The sample system should

be designed to avoid any flow fluctuation du

ri

ng measurement.

Use a flowmeter to observe the flow reading as shown in the

external piping diagrams of Figures 2.5, 2.6 and 2.7.

2.5.5 Preparation of Calibration Gas

Routine calibration using calibration gases is required for

optimizing the performance of this instrument. Once a week is a
suggested calibration frequency. Table 2-2 indicates the zero and span
gas required for calibration.

Table 2-2: Zero and Span Gas

Analyzer without

O

2

Analyzer with built-in

O

2

Analyzer with external

zirconia O

2

sensor

Zero gas

N

2

gas

N

2

gas

Dry air

Span gas
other than for
O

2

t

Gas concentration
of 90% or more of
full scale

Gas concentration of
90% or more of full
scale

Gas concentration of 90%
or more of full scale

Span gas for
O

2

analysis

NA

Gas concentration of
90% or more of full
scale

1 to 2% O

2

Advertising