Teledyne 7600 - Infrared gas analyzer - operating manual User Manual

Page 49

Advertising
background image

Infrared Gas Analyzer

Operation

Teledyne

Analytical

Instruments

49

This brings up the Cal. Settings ZERO Cal. screen where you can

set whether all components or individually selected components are
zeroed during a manual zero calibration. The choices are:

• at once—these components will be zeroed simultaneously

during a manual zero.

• each—these components will be zeroed individually during a

manual zero.

To change the manual zero setting:

1. In the “Cal. Settings ZERO Cal.” screen, point the cursor

to the channel you want to set by using the UP or DOWN
key. Press the ENT key.

2. Select “at once” or “each” by pressing the UP or DOWN

key to toggle between the two options. Press the ENT
key when the desired option is displayed.

To close the Cal. Settings ZERO Cal. screen or cancel this mode

midway, press the ESC key. The previous screen will return.

Example:

The options “each” or “at once” can be determined for each

channel.

• When the setting is “each”, select the channel on the manual

zero calibration screen (see Section 3.7) and perform a zero
calibration. Only that component is zeroed.

• When the setting is “at once”, performing a manual

calibration (see Section 3.7) will zero calibrate all
components tagged “at once”. You will notice on the manual