Section 6: Service & Maintenance

TROUBLESHOOTING

Important Precaution

Before handling the FB3000 boards or other system
components, a service technician must always be properly
grounded.

Electrostatic Discharge (ESD) can easily damage the
FB3000 board assemblies.

Remove the anti-static packing only when the parts are
ready for installation.

Handle the boards by their edges, and avoid touching their
components.


Error Condition(s)

Solution(s)

Check that scale is empty.
If Scale is empty, Call for

Service.

Load Cell(s) bad.

• A large amount of weight is zeroed.
• This

is

normal.

• Press

OK

and continue weighing.

• Possible load cell damage.
• Call for Service.

Load Cell Failure(s)
Flashing and displays “

- - - -

• Possible load cell damage.
• Call for Service.
• Access

the

Load Cell Diagnostics Menu

to verify the

load cell status.

• Count stability or change of counts.
• Contact the local service for further trouble-shooting.

SC Cells Found None

• Possible damaged load cell cable.
• Load

cell

shortened.

• Defective Pit Power Supply.
• Defective Smart Sectional Controller(s).
• Defective Analog Assembly.

Displays “

- - - -

“ ~ lb GROSS

• Communication error to load Cells.
• Check settings by pressing

F10

.

• Settings should be COM2, Even.