2.2.7. Installation Considerations

Thorough application review should be conducted before, during and after installation:

Determining Omnicell Capacity - Sizing the Omnicell (general rule) – Take
the overall live and dead load of the vessel. Multiply by 2, then divide by the
number of support points (legs, typically). For example:

4 Legs
26,000 lbs – dead load
74,000 lbs – live load

26,000

+ 74,000

100,000

x 2

200,000

÷ 4

50,000 lb for

each Omnicell

Loading – The vessel will have 3, or more, tank legs. The Omnicells should be
level and, under extreme cases, shimmed, so that the load is roughly equal on
each Omnicell.

Flooring - A thorough inspection of the flooring’s deflections must be considered
before installing an Omnicell system. Flooring must be rigid enough to handle the
total load without deflection.

Thermal contraction and expansion – Horizontal vessels, in direct sunlight,
can create thermal expansion issues. The 9104 FB Series Omnicell can handle
most thermal expansion/contraction issues.

Uplift and lateral constraints – In extreme conditions (high winds or seismic
activity) uplift and lateral stresses on the Omnicell system are a concern and
should be considered when choosing the best Omnicell for the application. The
9104 FB Series Omnicell provides 100% uplift and lateral checking, refer to the
product QuickFacts for a detailed explanation of the forces.

Isolate vessel – Many external factors affect the accuracy and stability of
an Omnicell system. Load cell signals can be affected by EMF sources such as
electric forklifts, pumps, motors, etc. A well-isolated tank will provide more
reliable results with fewer problems.