4 ph-b series specifications – Gentec-EO PH User Manual

Page 10

Advertising
background image

Version 2.3

PH Series Instruction Manual

Gentec Electro-Optics Inc. All rights reserved

6

Minimum Power

a

w/ OD-2

60 nW @ 980 nm

60 nW @ 980 nm

60 nW @ 850nm

200 nW @ 1550 nm

Spectral range
w/ OD-2

630 to 1080 nm

630 to 1100 nm

630 to 1080 nm

950 to 1650 nm

Uncertainty w/ OD-0.3,
1 or OD-2 attenuator

420-980 nm: ±4%

981-1049 nm: ±5%

1050-1080 nm: ±7%

400-1049 nm: ±5%

1050-1100 nm: ±7%

210-1049 nm: ±5

1050-1080 nm: ±7%

±5%


Specifications subject to change without notice.

1.4 PH-B SERIES SPECIFICATIONS

Table V. PH-B Series Specifications for Gentec-eo Monitors

Detector

PH10B-Si

PH5B-Ge

Monitor Compatibility

Maestro, M-Link, S-Link

Absorber

Silicon UV

Germanium

Spectral range

210 to 1080 nm

800 to 1650 nm

Peak sensitivity

980nm

1550nm

Sensitivity, typical

15 kV/W @ 633nm

80 kV/W @ 1047 nm

Maximum measurable
power

M-Link : 200 µW

S-Link : 175 µW

Maestro : 150 µW

(@ 633nm)

M-Link : 40 µW

S-Link : 30 µW

Maestro : 25 µW

(@ 1310 nm)

Maximum Average
Power Density

65 mW/cm

2

(@ 532 nm)

<320 mW/cm

2

(@ 1064 nm)

Noise equivalent power

50 pW (@ 633nm)

40 pW (@ 1310nm)

Minimum Power

e

1.5 nW (@ 633nm)

1.2 nW (@ 1310nm)

Damage threshold

[Maximum Average Power
Density]

100 W/cm

2

100 W/cm

2

Uncertainty

210 - 219 nm: ±8.0%
220 - 399 nm: ±6.5%
400 - 899 nm: ±2.5%
900 - 999 nm: ±3.5%

1000-1049 nm: ±5.0%
1050-1080 nm: ±7.0%

800- 1650 nm : ±3.5%

Aperture

10 mm diameter

5 mm diameter

e

Warm-up until the reading without laser power is stable for several minutes before offset nulling (zeroing the

display). Null the offset after each new power up. Half an hour warm-up is recommended for measuring low powers.
Temperature ±0.5 degrees.display). Null the offset after each new power up. Half an hour warm-up is
recommended for measuring low powers. Temperature ±0.5 degrees. Minimum Power = 30 X NEP