1 ph series photo detectors, 1 introduction – Gentec-EO PH User Manual

Page 5

Advertising
background image

Version 2.3

PH Series Instruction Manual

Gentec Electro-Optics Inc. All rights reserved

1

1 PH SERIES PHOTO DETECTORS

1.1 INTRODUCTION

The Gentec-EO Photo Detector family includes nine photo detectors sensors..

The PH100-Si, PH100Si-HA detectors use a silicon photodiode.
The PH100-SiUV, PH10B-Si, PE10B-Si, PE3B-Si detectors also use a silicon photodiode but have
enhanced sensitivity at shorter wavelengths.
The PH20-Ge, PH5B-Ge, PE5B-Ge detectors use a germanium photodiode.
The PE3B-In detectors use a InGaAs photodiode.

All detector heads are 27.4mm thick by 38.1 mm in diameter.

Table I. Measuring ranges of Gentec-EO PH photo detectors

Configuration

PH100-Si series

PH100-SiUV series

PH20 series

Detector alone

0.3 nW to 30 mW

0.3 nW to 4 mW

2 nW to 30 mW

With OD-0.3 attenuator

With OD-1 attenuator

-

3 nW to 300 mW

0.6 nW to 8 mW

3 nW to 38 mW

-

20 nW to 300 mW

With OD-2 Attenuator

30 nW to 750 mW

30 nW to 30 mW

200 nW to 500 mW

Note: quoted maximum ranges are average power at the 1064 nm wavelength for PH100-Si and PH20-Ge, 532 nm for PH100-

SiUV, and 850 nm for PH100-SiUV with OD-2

quoted minimum ranges are average power at the 980nm for PH100-Si, 850 nm for PH-100-SiUV and 1550 nm for PH20-

Ge (Minimum ranges = 30 X NEP)


Table II. Measuring ranges of Gentec-EO PH-B and PE-B photo detectors

PH10B-Si

1.5nW to 200µW

PE10B-Si

1.5 pJ to 0.15µJ

PH5B-Ge

1.2nW to 40µW

PE5B-Ge

PE3B-Si

PE3B-In

500 fJ to 3.0nJ

8 fJ to 30pJ

15 fJ to 300pJ

Note PH : quoted minimum and maximum ranges are average power at the 633nm for PH10-Si and 1310nm for PH5B-Ge

(Minimum ranges = 30 X NEP). The M-Link was used for those ranges.

Note PH : quoted minimum and maximum ranges are average power at the 633nm for PE10/3B-Si and 1310nm for PE3/5B-Ge/In

(Minimum ranges = NEE). The M-Link was used for those ranges.


The PH and PE series are supplied with a 180 cm long flexible cable terminated with a DB-15
"intelligent" male connector, for use with Gentec-EO Monitors.

NOTE: To eliminate possible damage, do not carry the detector using the connector cable.


The PH and PE photo detectors may also be supplied with a stand.

Advertising