3 ph series specifications – Gentec-EO PH User Manual

Page 8

Advertising
background image

Version 2.3

PH Series Instruction Manual

Gentec Electro-Optics Inc. All rights reserved

4

1.3 PH SERIES SPECIFICATIONS

Table IV. PH Series Specifications for Gentec-eo Monitors

Detector

PH100-Si-HA

PH100-Si

PH100-SiUV

PH20-Ge

Monitor Compatibility

Maestro, M-Link, P-Link, Uno, Tuner, Solo 2, Solo PE

Absorber

Silicon

Silicon

Silicon UV

Germanium

Bandwidth

(20 mA to 1 mA) 31 kHz
(1 mA to 20

A) 31 kHz

(20

A to 2

A) 31 kHz

(2

A to 0.1

A) 29 Hz

(0.1

A to 1 nA) 16 Hz

(Bandwidth Sampling Rate 50kHz, 2048 average values.)

Integra :

(25 mA to 1 uA) 32 kHz

( 1 uA to 1 nA) 340 Hz

Secondary stage: 3 Hz

Minimum repetition
rate for average power
measurements of a
pulse laser

(20 mA to 1 mA) 155 kHz
(1 mA to 20

A) 155 kHz

(20

A to 2

A) 155 kHz

(2

A to 0.1

A) 145 Hz

(0.1

A to 1 nA) 80 Hz

Integra :

(25 mA to 1 uA) 155 kHz

( 1 uA to 1 nA) 1700 Hz

Spectral range

350 to 1080 nm

320 to 1100 nm

210 to 1080 nm

800 to 1650 nm

Peak sensitivity

980 nm

980 nm

980 nm

1550 nm

Maximum Power typ.

[@ Power Density]

36 mW

[36 mW/cm

2

]

(@ 1064 nm)

36 mW

[36 mW/cm

2

]

(@ 1064 nm)

4 mW

[65 mW/cm

2

]

(@ 532 nm)

30 mW

[< 320 mW/cm

2

]

(@ 1064 nm)

Minimum Power

a

0.3 nW @ 980 nm

0.3 nW @ 980 nm

0.3 nW @ 850 nm

2 nW

@ 1550 nm

Damage threshold

[Maximum Average
Power Density]

100 W/cm

2

100 W/cm

2

100 W/cm

2

100 W/cm

2

Typical Detector
saturation current

b

6.3 mA/cm

2

6.3 mA/cm

2

17.6 mA/cm

2

140mA/cm

2

a

Warm-up until the reading without laser power is stable for several minutes before offset nulling (zeroing the

display). Null the offset after each new power up. Half an hour warm-up is recommended for measuring low powers.
Temperature ±0.5 degrees.display). Null the offset after each new power up. Half an hour warm-up is
recommended for measuring low powers. Temperature ±0.5 degrees. Minimum Power = 30 X NEP

Advertising