1 example – Gentec-EO PH User Manual

Page 17

Advertising
background image

Version 2.3

PH Series Instruction Manual

Gentec Electro-Optics Inc. All rights reserved

13

Slope: The slope of the linear interpolation.

Sens

LOW_

λ

: Sensitivity at

LOW

.

Sens

HIGH_

λ

: Sensitivity at

HIGH

.

LOW

: The next lowest available wavelength near your desired wavelength.

HIGH

: The next highest available wavelength near your desired wavelength.

DESIRED

: Desired Wavelength.

: The difference between the desired wavelength and the inferior wavelength.

4.5.1 Example


You have a PH100-Si and your laser is at 632.8 nm.

See your CERTIFICATE OF CALIBRATION for the sensitivity of your power detector as a function of
the wavelength.

Wavelength (nm)

Sensitivity (A/W)

620

0.32

630

0.35

640

0.37

650

0.40

660

0.43


632.8 nm is between 630 nm and 640 nm therefore,

Sens

LOW_

λ

= 0.35 A/W

Sens

HIGH_

λ

= 0.37 A/W

LOW

= 630 nm

HIGH

= 640 nm

: 632.8 - 630 = 2.8 nm

Slope: (0.37-0.35)/(640-630) = 0.002

Sensitivity

desired_

: 0.35 + 2.8 * 0.002= 0.356 A/W


Advertising