1.5 PE-B SERIES SPECIFICATIONS

Table VI. PE-B Series Specifications for Gentec-eo Monitors

Detector

PE10B-Si

PE5B-Ge

PE3B-Si

PE3B-In

Monitor Compatibility

Maestro, M-Link, S-Link

Absorber

Silicon UV

Germanium

Silicon UV

InGaAs

Spectral range

210 to 1080 nm

800 to 1650 nm

210 to 1080 nm

f

900 to 1700 nm

f

Sensitivity, typical

20 MV/J @ 634 nm

1 GV/J @ 1310 nm

100 GV/J @ 634 nm

10 GV/J @ 1310 nm

Rise Time (0-100%)

30 µs

25 µs

15 µs

12 µs

Max Repetition Rate

1000 Hz

1000 Hz

1000 Hz

1000 Hz

Max Pulse Width

10 µs

10 µs

10 µs

10 µs

Maximum

Measurable Energy

M-Link : 0.15 µJ

S-Link : 0.13 µJ

Maestro : 0.11 µJ

(@ 634nm)

M-Link : 3.0 nJ

S-Link : 2.5 nJ

Maestro : 2.0 nJ

(@ 1310 nm)

M-Link : 30 pJ

S-Link : 25 pJ

Maestro : 20 pJ

(@ 634 nm)

M-Link : 300 pJ

S-Link : 250 pJ

Maestro : 200 pJ

(@ 1310 nm)

Maximum Energy

Density

5 µJ/cm

2

5 µJ/cm

2

n/a

n/a

Max Average Power

Density (@ 1064nm)

65 mW/cm

2

(@ 532 nm)

<320 mW/cm

2

(@ 1064 nm)

n/a

n/a

Noise Equivalent

Energy

1.5 pJ @ 634nm

1 pJ @ 1310 nm

8 fJ @ 634nm

30 fJ@ 1310nm

Uncertainty

210 - 219 nm: ±8.0%
220 - 399 nm: ±6.5%
400 - 899 nm: ±2.5%
900 - 999 nm: ±3.5%

1000 -1049 nm:

±5.0%

1050-1080 nm:

±7.0%

800-1650 nm: ±3.5%

±4% only @ 634 nm

±4% only @ 1310 nm

Effective Aperture

10 mm Ø

5 mm Ø

3 mm Ø

3 mm Ø

Active Area

0.9 cm

2

0.2 cm

2

0.07 cm

2

0.07 cm

2

Dimensions

38.1mm Ø x 27.4Dmm

Weight

91 g

Beam Position

Dependence

±1% @ 652 nm

±3% @ 1064 nm

±1% @ 1064 nm

±3% @ 800 nm

n/a

n/a


2 OPERATING INSTRUCTIONS

f

The photodiode is calibrated at one wavelength and spectrally correted according to a typical

curve.