Calibration and setting flow – SKC 228-9530 QuickTake 30 User Manual

Page 14

Advertising
background image

13

Tubing

Calibrator outlet

Pump inlet

Spore trap cassette

Note: Use a second cassette adapter to

attach calibrator to a spore trap cassette.

Digital display:

Calibration mode

Calibration with a spore trap cassette

Cassette adapter

Digital display:

Store setting

Calibration and Setting Flow

Use a primary standard calibrator to calibrate pump fl ow rate.

Calibrate the fl ow rate with a representative sampling medium in line.

Before use, allow the pump to equilibrate aft er moving it from one temperature
extreme to another.

1. Ensure the pump has run for 5 minutes before calibrating. Using 3/8-inch

ID fl exible tubing and appropriate adapters, connect the inlet of the pump
to the outlet of a representative sampling medium and a primary standard
calibrator to the inlet of the representative sampling medium (as shown
above).

2. Scroll to any run time preset. Press and hold the CAL

butt on for two seconds. The pump will start running
and a fl ashing CAL will appear on the LED.

3. Press  or  to increase or decrease the fl ow until

the desired fl ow rate is displayed on the calibrator.

4. Press

. A fl ashing Stor will appear on the LED.

To save new sett ing: Press [



]. The pump

will stop running.

To ignore new sett ing: Press

. The pump will

stop running.

Note:

A security code is not needed to change fl ow

rate. It is only required when changing factory
sett ings. To change factory sett ings, see Advanced
Setup.