Quicktake 30 quick guide – SKC 228-9530 QuickTake 30 User Manual

Page 3

Advertising
background image

1

QuickTake 30 Quick Guide

Terms »

Star Button

Turns on LED, toggles between displays, and exits without saving

Up and Down Arrow Buttons 

Scroll through presets, increase or decrease sampling parameters, and toggle

between setup options

Button Sequence

or  = Press individually

[] = Press both simultaneously



= Security code, press in sequence

Programming Sequences »

• To activate pump LED:

Press and hold  for 2 seconds.

• To check battery status:

Press .

• To choose a run time preset:

Press  or  to scroll through presets. Once desired preset displays, press

[] to start sampling.

• To Run or Hold:

Press [].

• To repeat a sample run:

From Done, reset (rSEt), or sampling error (SErr), press [] to return to run

time preset. Press [] to sample.

• To change or calibrate fl ow:

Note: Flow rate does not appear on LED.

Press  or  and scroll to a run time preset. Press and hold CAL for 2 seconds.

(Does not require a security code.) Press  or  to change fl ow rate until desired

fl ow appears on calibrator. Press . When Stor displays, press [] to save

setting or  to ignore changes to fl ow setting.

• To interrupt run, terminate a sample, or reset pump:

While in Hold, press . Sample reset (rSEt) will display. Press [] to reset

pump or  to ignore reset and return to Hold.