Setup, Timed run – SKC 228-9530 QuickTake 30 User Manual

Page 15

Advertising
background image

14

Setup

Press  or  to scroll through the run time presets on the LED.

Run Time Mode

LED Display

Timed Run (defaults: 15, 10, 5, 2, and 1 min)

t xx

Intermitt ent Sampling

Intt

Continuous

Run

Cont

The user may change run time presets. See Advanced Setup, Programming Your Own
Run Time Presets.

Timed Run

Run time accumulates only while the pump is running.

Timed Run mode is designed to perform one sample for a
predetermined time from 1 to 999 minutes (selected from
presets). The user starts the pump running and the pump
automatically stops running after the sample time has
elapsed.

1. Press  or  to scroll to the desired run time preset.
2. Press [



] to select the desired preset. The pump will

start running and the LED will display a count down to
zero (run time remaining below one minute is displayed
in seconds with a colon). The pump will stop and beep
four times. The display will fl ash Done.

3. Press

to return to the run time preset display. If the

user wishes to repeat the sample, press [



].

User Options During Timed Run Sampling
Viewing batt ery status

Press

. Display will automatically return to run time

remaining aft er fi ve seconds or press

again.

Hold mode

Press [



] while the sample is running to place the

pump and timer in Hold. Run time remaining and Hold
will display alternately. Press [



] while in Hold to

continue the sample run.

Terminating a sample and resett ing the pump

Press

while in Hold. Sample Reset (rSEt) will fl ash on

the LED.
To terminate the sample and reset the pump: Press [



].

To continue the sample run: Press

to return the display to Hold. Press

[



].

Digital display:

Timed Run 2 minutes

Digital display:

Run time remaining is

2 seconds

:

Digital display:

Sample run completed