Performance profi le – SKC 228-9530 QuickTake 30 User Manual

Page 5

Advertising
background image

4

Performance Profi le

Flow Range:

10 to 30 L/min

Flow Compensation
Accuracy:

± 5% of set fl ow

Compensating Flow

10 L/min at 90 inches water back pressure

Back Pressure Range: 20 L/min at 50 inches water back pressure

30 L/min at 15 inches water back pressure

Power:

Rechargeable lithium-ion (Li-Ion) battery, 7.4 V, 9.6-Ah

capacity, 71.04 Wh

Battery with AC charger/adapter, 100-240 V

Battery Recharge Time: Approximately 5 hrs

(varies with batt ery capacity
and level of discharge)

Typical Cumulative

Spore Trap* (e.g., VersaTrap): 5 hrs at 15 L/min (battery only)

Run Time:

BioStage viable cascade impactor: 4 hrs at 28.3 L/min (battery only)

25-mm, 1.2-μm MCE fi lter: 9+ hrs at 10 L/min

To achieve, use QuickTake 30 with a fully

charged battery and AC charger/adapter.

37-mm, 0.8-μm MCE fi lter: > 14 hrs at 10 L/min (battery only)

* Sampling times when using spore traps are usually ≤ 10 min. SKC

recommends reducing length of sample time when using a 30 L/min
fl ow rate to prevent overloading the media.

† Results obtained using a new pump and new fully charged batt eries. Pump and

batt ery performance may vary.

The AC charger/adapter can be used to extend batt ery run
time, but it is not a batt ery eliminator. Therefore, it will
not provide indefi nite run times.

Temperature:

Operating:

32 to 113 F (0 to 45 C)

Storage:

-4 to 95 F (-20 to 35 C)

Charging:

32 to 113 F (0 to 45 C)

Operating Humidity:

0 to 95% non-condensing

Protect sample pump from weather when in use outdoors.

Compare the information in this table to pump compensation range to determine appropriate applications.

Typical Back Pressure of Sampling Media (inches water)

Flow Rate (L/min)

10

12

15

20

Filter/Pore Size (μm)
25-mm MCE, 0.8

65

80

103

148

25-mm MCE, 0.45

138

172

> 200

37-mm MCE, 0.8

22

28

36

51

37-mm PVC, 5.0

9

11

15

21