TDR module 1 hardware Skew control
Skew1
From -5e-011 to 5e-011 step 1e-013
Double
' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")
' Change TDR Pod 1 skew to -15pS
app.TDR.Skew1 = -15e-12
Range
Example
Description
TDR Edge polarity for TDR2 trace
Polarity2
Enum
' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")
' Change TDR trace 2 polarity to negative
app.TDR.Polarity2 = "Neg"
Example
Description
Values
Neg
Pos
TDR module 2 hardware Skew control
Skew2
From -5e-011 to 5e-011 step 1e-013
Double
' Visual Basic Script
Set app = CreateObject("LeCroy.XStreamDSO")
' Change TDR Pod 2 skew to -15pS
app.TDR.Skew2 = -15e-12