Chapter four – Teledyne LeCroy WaveExpert series Automation Manual User Manual
Page 212

Chapter Four
Control Reference
Sets/Queries the maximum number of values from the source result to include in the histogram
function Fx.
Values
From 20 to 1000 step 1
Integer
Range
Description
Sets/Queries the number of bins in the histogram function Fx.
Bins
From 20 to 2000 step 1, locked to 1 2 5, fine grain allowed=false, on=false
DoubleLockstep
Range
Description
Sets/Queries the horizontal scale in units per division for the histogram function Fx. Use the FindScale
control to automatically determine the scale by looking at the non-zero populated bins.
HorScale
From 1e-012 to 1e+012 step 0.01, locked to 1 2 5, fine grain allowed=false, on=false
DoubleLockstep
Range
Description
Sets/Queries the horizontal value at the center of the graticule of the histogram function Fx.
Center
From -1e+010 to 1e+010 step 1e-012
Double
Range
Description
Creates a suitable horizontal position and scale to include all the non-empty bins of the histogram Fx.
FindScale
Action
Description
916435 RevA
4-8