3.5.3. Performing a Quick Test

The ATRT provides a Quick Test mode that can be used to measure a transformer’s turns ratio
by only pressing a single button. To initiate a Quick Test:

a. Start from the “START-UP” menu:

Press and hold down the

[START]

key.

b. The following screen will be displayed:

Continue to hold down the

[START]

key until the unit beeps again, and then release

the

[START]

key.

c. The following screen will be displayed while the test is being performed:

The test results will be displayed on the LCD screen when testing has finished:

Press the

[STOP]

key to return to the “START-UP” menu.

RATIO mA
+1.003 0002

SINGLE PHASE XFORMER

PLEASE WAIT

TEST IN PROGRESS

“QUICK TEST”

HOLD START KEY UNTIL

BEEP FOR QUICK TEST

1.TEST XFMR 06/02/10
2.SETUP 15:33:10
3.CALCULATOR