REV 3 EZCT S2 USER’S MANUAL
51
5.2
Verifying the EZCT S2’s V
x
Sense Circuit Using an External Meter
The excitation voltage (V
x
) sensed by the EZCT S2 can be verified using an external RMS volt
meter. Follow the steps below to verify the EZCT S2’s V
x
sense circuit:
a. Connect the X cables to an RMS volt meter as shown in Figure 13.
b. Select the EZCT S2’s Diagnostics mode (see section 5.1).
c. Press the
[PAPER ∧ Contrast]
key to raise the V
x
voltage.
d. Verify the V
x
voltage displayed on the EZCT S2 LCD screen against the value displayed on
the external volt meter.
e. Press the
[STOP]
key to end the test.
Figure 13. EZCT S2 V
x
Verification Test Connections