ZOLL Base PowerCharger

1x1

Operator’s Manual

18

*XLGDQFH DQG 0DQXIDFWXUHU¶V 'HFODUDWLRQ  (OHFWURPDJQHWLF ,PPXQLW\

The ZOLL Base PowerCharger

1x1

is intended for use in the electromagnetic environment specified below. The customer or

the user of the ZOLL Base PowerCharger

1x1

should ensure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment - guidance

Electrostatic
discharge (ESD)
IEC 61000-4-2

± 6 kV contact
± 8 kV air

± 6 kV contact
± 8 kV air

Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.

Electrical fast
transient/burst IEC
61000-4-4

± 2 kV for power
supply lines
± 1 kV for input/output
lines

± 2 kV for power
supply lines
± 1 kV I/O

Mains power quality should be that of a typical
commercial or hospital environment.

Surge
IEC 61000-4-5

± 1 kV differential
mode
+/- 2 kV common
mode

± 1 kV differential
mode
+/- 2 kV common
mode

Mains power quality should be that of a typical
commercial or hospital environment.

Voltage dips, short
interruptions and
voltage variations
on power supply
input lines.
IEC 61000-4-11

<5% Ut (>95% dip in
Ut) for 0.5 cycle
40% Ut (60% dip in
Ut) for 5 cycles
70% Ut (30% dip in
Ut) for 25 cycles
<5% Ut (>95%dip in
Ut) for 5 sec

<5% Ut (>95% dip in
Ut) for 0.5 cycle
40% Ut (60% dip in
Ut) for 5 cycles
70% Ut (30% dip in
Ut) for 25 cycles
<5% Ut (>95%dip in
Ut) for 5 sec

Mains power quality should be that of a typical
commercial or hospital environment. If the user of
the ZOLL Base PowerCharger

1x1

requires

continued operation during mains interrupting, it
is recommended that the ZOLL Base
PowerCharger

1x1

be powered from an uninterruptible

power supply.

Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment

1RWH

8W LV WKH DF PDLQV YROWDJH SULRU WR DSSOLFDWLRQ RI WKH WHVW OHYHO