ZOLL Base PowerCharger

1x1

Operator’s Manual

4

6SHFLILFDWLRQV

%DVH 3RZHU&KDUJHU

[

*HQHUDO

'LVSOD\V

&RQWUROV

=2// %DWWHULHV

Size

8.2 cm H x 15.8 cm W x 31.8 cm L
(3.2 in. x 6.2 in. x 12.5 in.)

Weight

2.1 kg (4.5lbs) without batteries.

Equipment
Classification

Class 1, No Patient Applied Parts,
Ordinary, Portable, Continuous Operation

Input Rating

100-120 VAC, 50/60 Hz, 340mA. (or)
220-240 VAC, 50/60 Hz, 170mA.

Safety
Standards

Designed to meet or exceed:
EN60601-1 Medical Electrical Equipment
Part 1: General Requirements for Safety.

Designed to meet or exceed:
IEC 60601-1-2 Medical Electrical
Equipment Part 1-2: General
Requirements for Safety.
Collateral standard:
Electromagnetic Compatibility -
Requirements and Tests.

UL2601-1 Standard for Medical Electrical
Equipment. Part 1: General Requirements
for Safety with respect to Electrical Shock,
Fire and Mechanical Hazards only.
(UL Classification mark associated with
100-120 VAC input only.)

Recharge Time

QuickCharge Four hours or less for total recharge of fully

depleted battery for PD 4410/Smart
Battery. Eight hours or less for XL Smart
Battery.

AutoTest

Eight hours or less for AutoTest and total
recharge of battery for PD 4410/Smart
Battery. Fifteen hours or less for XL Smart
Battery.

Temperature

5 °C to 40 °C(operating)
15°C to 35°C (optimal battery charging)
-40°C to 70°C (storage and transportation)

Humidity

5% to 95% (relative humidity - non
condensing).

External Fuses (2 ) 400 mA/250V Time Delay fuses for

100-120 VAC input.
(2) 200mAL/250V, Type T fuses for
220-240 VAC input.

POWER

I

lluminates when Base PowerCharger

1x1

is

plugged into live AC mains.

CHARGER ON

Indicates battery is being charged.

BATT. READY

Illuminates when the charge cycle is
complete.

FAULT

Illuminates when battery capacity test or
battery compartment test fails

TESTING

Illuminates when either a battery test or a
battery compartment test is in progress.
Base PowerCharger

1x1

with AutoTest only:

Illuminates with the CHARGER ON light
during AutoTest and recharge of the battery.

TEST Button

Illuminates TESTING indicator light and
initiates battery test cycle or battery
compartment test.
Automatically charges, discharges and
recharges the battery.

Type

Rechargeable , sealed lead acid.
ZOLL Battery Pack PD 4410
ZOLL Smart Battery
ZOLL XL Smart Battery

Weight

PD 4410/Smart Battery: 1 kg (2.2 lbs)
XL Smart Battery: 1.68 kg (3.7 lbs)

Size

PD 4410/Smart Batttery:

4 cm H x 7 cm W x 18.9 cm L
(1.58 in. x 2.78 in. x 7.44 in.)

XL Smart Battery:

9.4 cm H x 7 cm W x 20 cm L
(3.7 in. x 2.78 in. x 7.9 in.)

Capacity

PD 4410/Smart Batttery:

2.5 Amp Hour rating for new batteries

XL Smart Battery:

4.5 Amp Hour rating for new batteries

Nominal Battery
Voltage

2.0 V/cell; 5 cells.

Temperature

0 °C to 50 °C(operating)
15°C to 35°C (optimal battery charging)
-40°C to 70°C (storage and transportation)

Humidity

5% to 95% (relative humidity - non
condensing).

Capacity Test

At a minimum, test batteries every three
months. Refer to ZOLL’s Battery
Management Program brochure.