0dqxdo %dwwhu\ 7hvwlqj 3urfhgxuh, Operation – ZOLL M Series Defibrillator Rev B Base PowerCharger 1x1 User Manual

Page 13

Advertising
background image

9

Operation

2.

Insert battery into the compartment. Make sure the
battery is fully seated (A click can be heard).

3. The CHARGER ON and TESTING indicators illuminate

and remain on during complete AutoTest and recharge
cycle (shown). Only the CHARGER ON indicator
illuminates during a QuickCharge recharge cycle.

4. In 8 hours or less for the PD 4410/Smart Battery or 15

hours or less for the XL Smart Battery (using AutoTest)
the CHARGER ON light will go off and the BATT.
READY
light will illuminate indicating the battery has
passed the capacity test and is fully charged and ready
for use.
In 4 hours or less for the PD 4410/Smart Battery or 8
hours or less for the XL Smart Battery (using
QuickCharge) the CHARGER ON light will go off and the
BATT. READY light will illuminate indicating the battery
has been charged to 100% of its present capacity and is
ready for use.

If the battery or charger is not capable of charging, the
FAULT light will illuminate and the CHARGER ON light
may stay on. Before discarding the battery, test the
battery compartment for defective operation (see
Operational Checks Section) and perform a second
battery test.

5. To remove the battery, insert a finger into the recess at

the top of the battery compartment, press against the
battery to disengage the battery locking clip and lift the
battery out. A charger test is performed illuminating all
LED's briefly followed by the BATT. READY and
TESTING indicators illuminating for 2 seconds and then
all LED's go off except for the POWER LED.

0DQXDO %DWWHU\ 7HVWLQJ 3URFHGXUH

Although this test is available, the Auto Test feature
eliminates the need to manually test batteries since battery
capacity is automatically tested with each battery recharge
(except when a fully charged battery is placed in the unit).
It is not necessary to insert a fully charged battery for
testing; the Base PowerCharger

1x1

will automatically test the

battery regardless of charge status.
To test a battery:
1.

Check to see that the Base PowerCharger

1x1

is con-

nected to live AC mains. The POWER light should be
lit.

2.

,QVHUW EDWWHU\ LQWR WKH FKDUJHU 0DNH VXUH WKH EDWWHU\

LV IXOO\ VHDWHG (A click can be heard).

3. Press the TEST button. The TESTING indicator light will

illuminate for the duration of the test cycle.

4. If the battery passes the test, the BATT. READY light will

illuminate while the TESTING light remains lit indicating
the battery is fully charged and ready to use. See the M
Series Operator's Manual for run time information.

5. If a battery fails the test, the FAULT light will illuminate

while the TESTING light remains lit. Before discarding
the battery, test the battery compartment for possible
defective operation (See Operational Checks Section).
Repeat Battery test.

6. Record the date and results of the test on the battery

label.