General Information

Warnings and Cautions

:DUQLQJV
*HQHUDO

The Base PowerCharger

1x1

has been tested against

interference from radio frequency emissions typical
of two-way radios used in emergency service/public
safety activities. Users of the Base
PowerCharger

1x1

should refer to Section 5 to

determine recommended operation distances from
RF transmission equip

ment.

Use only the AC power line cord supplied with the
device. Failure to use the proper line cord could
result in excess leakage currents and reduced
safety.

Do not disassemble the Base PowerCharger

1x1

. A

shock hazard exists. Refer all problems to ZOLL
Technical Service.

Follow all recommended maintenance instructions.
If a problem occurs, obtain service immediately. Do
not use the Base PowerCharger

1x1

until the unit has

been inspected by the appropriate personnel.

Do not place anything on top of or beneath the Base
PowerCharger

1x1

, such as blankets or cloths. Doing

so may block the vents on the unit, preventing
proper dissipation of heat during operation.

Do not use adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary
the Base PowerCharger

1x1

should be observed to

verify normal operation.

Do not sterilize the Base PowerCharger

1x1

.

Do not use alcohol or ketones (MEK, acetone, etc.)
on the Base PowerCharger

1x1

.

Always wait until the Base PowerCharger

1x1

has

completed its charge cycle before removing
batteries.

Repeated use of partially charged batteries will
reduce their capacity and operating life.

Do not use a battery unless the BATT. READY
indicator is lit and the FAULT light is off. Failure to do
so may result in the use of a fully depleted battery.

Do not install batteries into Monitor/Defibrillators
when storage may exceed 90 days. Battery damage
may occur.

2SHUDWRU 6DIHW\

Do not operate the Base PowerCharger

1x1

in the

presence of flammable agents (such as gasoline),
oxygen-rich atmospheres, nitrous oxide, or
flammable anesthetics. Using the device near the
site of a gasoline spill may cause an explosion.

Do not immerse or set the Base

3RZHU&KDUJHU

[

in

liquid. Using the device near or within puddles of
water may present a shock hazard to the operator,
patient, and nearby personnel.

3DWLHQW 6DIHW\

Do not mount the device directly above the patient.
Locate the Base

3RZHU&KDUJHU

[

and ZOLL

resuscitation devices where they cannot harm the
patient if they fall.

&DXWLRQ

When a fully charged battery is placed into the Base
PowerCharger

1x1

with the AutoTest, the AutoTest

will NOT activate and the BATT. READY indicator
will illuminate within a few minutes after inserting the
battery into the charger. Since the AutoTest
sequence is not performed, it is possible to have a
shorter battery run time than expected.