Troubleshooting

9650-1215-01 Rev. C

13

FAX ERROR message is displayed.

A dialing problem other than a busy line, no carrier, or no
connection occurred.

The E Series automatically re-transmits until the transmission is
successful or the E Series operator presses the Abort softkey.

FAX HANGUP message is displayed.

Receiving phone is not receiving transmission. The E Series
automatically attempts to transmit until the transmission is
successful or the E Series operator presses the Abort softkey.

FAX INTERRUPTED message is displayed.

Power cycle E Series unit and follow steps to attempt to
retransmit.

Contact ZOLL Technical Service department if problem persists.

FAX NO CARRIER message is displayed.

Receiving phone line is not picking up or the sending modem is
experiencing problems. The E Series automatically attempts to
transmit until the transmission is successful or the E Series operator
presses the Abort softkey.

If retransmission fails:

Check the modem connection to the E Series unit.

Power cycle the cellular phone.

FAX NO DIAL TONE message is displayed.

Check the modem connection to the E Series unit.

Check the modem connection to the phone jack or cellular phone.

Power cycle cellular phone.

MODEM REQUIRED message is displayed.

No PCMCIA modem card is present. Insert PCMCIA modem card
and repeat transmission attempt.

Symptom

Recommended Action

Advertising