ZOLL M Series Temperature

5

Printing Temperature Data

See the M Series Operator’s Guide Recorder Operation section for instructions on how to record data.
If you are currently displaying temperature, press the RECORDER button to print a strip chart that
includes the following values:

• T1
• T2

T

• Time of measurement.

Maintaining the Temperature Option

See the M Series Service Manual for instructions on maintaining the Temperature option.

Performing the Daily Checkout Procedure

The following is a checkout procedure that should be performed daily to ensure that the Temperature
option is functioning properly. In addition, this daily checkout procedure ensures that medical
personnel maintain familiarity with the proper use of the Temperature option. For more specific
instructions on how to accomplish each step listed below, refer to the related section in this manual.
Use only YSI 400-compatible temperature simulators. See "Ordering Accessories" on page 8 for a list
of ZOLL-approved temperature simulators. Daily checkout for the T2 channel also requires a ZOLL 2-
channel Y-adapter.

To perform the daily checkout procedure:

1. Connect the temperature simulator cable to the M Series unit and the temperature simulator.
2. Set the simulator to simulate 68

° F or 20° C.

3. Verify that the M Series displays 68

° F (+/- 0.2° F) or 20° C (+/- 0.2° C).

4. Set the simulator to simulate 104

° F or 40° C.

5. Verify that the M Series displays 104

° F (+/- 0.2° F) or 40° C (+/- 0.2° C).

Repeat steps 1 - 5 using the other temperature channel.