Applying a disposable sensor – ZOLL M Series Defibrillator Rev M User Manual

Page 10

Advertising
background image

M S

ERIES

O

PERATOR

S

G

UIDE

- O

PTION

I

NSERT

SpO

2

- 8

long, it may extend over and past the finger stop (see
Figure 3).

Figure 3

Note: With smaller digits, the digit may not need to be

pushed all the way to the stop to completely cover
the detector window.

3. Check the sensor position to ensure that the top and

bottom halves of the sensor are parallel. Complete
coverage of the detector window is needed to ensure
accurate data (see Figure 4).

Figure 4

Orient the sensor so that the cable runs towards the
top of the patient's hand.

4. Lift the clear plastic protective cover from the female

end of the patient cable, then plug the sensor cable’s
male connector all the way into the patient cable
connector (see Figure5).

Figure 5

5. Lower the clear plastic protective cover over the

connection to secure it (see Figure 6).

Figure 6

6. Connect the SpO

2

patient cable to the SpO

2

connector on the back of the M Series unit as shown
in Step 4 of “SpO2 Setup” on page 6.

Applying a Disposable Sensor

A disposable sensor may be reapplied to the same
patient if the emitter and detector windows are clear and
the adhesive still adheres to the skin.

The adhesive can be partially rejuvenated by wiping with
an alcohol wipe and allowing the sensor to thoroughly air
dry prior to replacement on the patient.

Note: LNCS Adtx sensors are not intended for use

across a child's hand or foot. For instructions on
proper application of neonatal sensors, refer to
the sensor’s Directions for Use.

1. Open the pouch and remove the sensor.

2. Holding the sensor with the tan printed side

downward, bend the sensor backward and remove
backing material from the sensor.

3. Orient the sensor so that the digit can be attached to

the detector side of the sensor first (see Figure 7).

Figure 7

4. Press the detector onto the fleshy part of the finger

near the tip of the finger. Complete coverage of the
detector window is needed to ensure accurate
measurements.

5. With the emitter positioned over the finger nail,

secure the wings around the finger (see Figure 8).

Figure 8

ZM 040022

Sensor Cable Connector

Patient Cable
Connector

Clear Protective
Cover

ZM 040024A

Detector Window

Advertising