Specifications, General – ZOLL M Series Defibrillator Rev M User Manual

Page 14

Advertising
background image

M S

ERIES

O

PERATOR

S

G

UIDE

- O

PTION

I

NSERT

SpO

2

- 12

Specifications

Note: The M Series Pulse Oximetry Option is calibrated for Functional saturation.

General

Saturation (% SpO

2

) Range

1% - 100%

Pulse Rate (bpm) Range

25 - 240 bpm

Saturation (% SpO

2

) Accuracy

During No Motion Conditions

Adults

70% - 100%, ±2 digits

0% - 69%, unspecified

Neonates

70% - 100%, ±2 digits

0% - 69%, unspecified

Saturation (% SpO

2

) Accuracy

During Motion Conditions

Adults

70% - 100%, ±3 digits

0% - 69%, unspecified

Pulse (bpm) Accuracy
During No Motion Conditions

25 - 240, ±3 digits

Pulse (bpm) Accuracy
During Motion Conditions

25 - 240, ±5 digits

SpO

2

Alarm Limits

On/Off displayed on monitor. User selectable.
High 72 - 100% saturation, Low 70 - 98% saturation

Pulse Rate Alarm Limits

On/Off displayed on monitor. User selectable.
High 60 - 235 bpm, Low 20 - 100 bpm

Saturation (% SpO

2

) Resolution

1%

Pulse Rate Resolution

1 bpm

Text Display Update Rate

100 milliseconds

Trace Update Rate

52 milliseconds

Bio-Compatibility

Patient contacting material meets requirements of ISO 10993-1,
Biological Evaluation of Medical Device - Part I, for external
devices, intact surfaces and short-term exposure.

Environmental

Operating Temperature:

0° to 40° C

Storage and Shipping Temperature: -20° to 60° C

Note: The M Series device may not perform to specifications

when stored at the upper or lower extreme limits of storage
temperature and immediately put into use

Electromagnetic Immunity
(SpO

2

Option Only)

AAMI DF-2: IEC 1000-4-3 to 18 v/m

Operating Time

For a new, fully charged battery pack at 20°C:

35 defibrillator discharges at maximum energy (360J), or

2.0 hours minimum of continuous ECG monitoring, or

1.75 hours of continuous ECG monitoring/pacing at 60 mA,
70 beats/min.

Advertising