Non-Invasive Blood Pressure
9650-000820-01 Rev. B
Propaq M Operator’s Guide
A-9
Non-Invasive Blood Pressure
Technique: Non-invasive oscillometric method
Operating Modes: Automatic and manual
Automatic Intervals: 1, 2, 3, 5, 10, 15, 30 and 60 minute intervals.
Turbocuf: Maximum measurements allowable in a 5 minute period
Pressure Measurement Range:
Systolic: 20 to 260 mmHg
Diastolic:10 to 220 mmHg
Mean: 13 to 230 mmHg
Static Pressure Accuracy: +/- 3 mmHg
Pulse Rate Range:
Adult:
30 to 200 +/- 5 BPM
Pediatric:30 to 200 +/- 5 BPM
Neonatal:35 to 220 +/- 5 BPM
Default Cuff Inflation Pressure:
Adult: 160 mmHg
Pediatric:120 mmHg
Neonatal:90 mmHg
Maximum Cuff Inflation Pressure:
Adult: 270 mmHg
Pediatric:170 mmHg
Neonatal:130 mmHg
Single Fault Backup Overpressure Limit:
Adult: 308 mmHg
Pediatric:205 mmHg
Neonatal:154 mmHg