Propaq M Operator’s Guide

6-1

Chapter 6

Monitoring ECG

This chapter describes how to use the Propaq M unit to monitor ECG.

You can use a 3-lead, 5-lead, or 12-Lead wire configuration for ECG monitoring (see Chapter
14, “12-Lead ECG Interpretive Analysis” for more information on 12-lead monitoring).

Warning!

•

Excessive body hair or wet, sweaty skin may interfere with electrode adhesion.
Remove the hair and/or moisture from the area where the electrode is to be attached.

•

Use only electrodes that are well within the expiration date indicated on the package.

•

Remove ECG electrodes from their sealed package immediately prior to use. Using
previously opened or out-of- date electrodes may degrade the ECG signal quality.

•

Monitoring electrodes may become polarized during defibrillator discharge, causing
the ECG waveform to briefly go off screen. ZOLL Medical Corporation recommends
the use of high quality silver/silver chloride (Ag/AgCl) electrodes to minimize this
effect; the circuitry in the instrument returns the trace to the monitor display within
a few seconds.

•

To assure protection against the effects of defibrillator discharge, use only
ZOLL-approved accessories.

•

To avoid a shock hazard and interference from nearby electrical equipment, keep
electrodes and patient cables away from grounded metal and other electrical
equipment.