Installation and User’s Guide

3

Verifying System Functionality

This completes the functional verification procedure.

15

For systems with the
Agilent 4284A precision
LCR meter,
open the
junction capacitance
model (juncap.mdl) in
IC-CAP.

a From the IC-CAP/Main window menu bar,

choose File > Examples... .

b In the Directories list of the File Open dialog

box, double-click on the directory
.../examples/model_files.

c In the Directories list of the File Open dialog

box, double-click on the directory
.../model_files/diode.

d In the Files list of the File Open dialog,

double-click on juncap.mdl.

•

This opens the File Open dialog box.

•

This opens a list of model files.

•

(Scroll the list, if needed.) This displays
a list of modeling files in the Files list of
the File Open dialog.

•

This opens the juncap model window.

•

The juncap model window contains tab
folders used to interact with the model
data.

16 Select the DUT/Setup.

a Select the DUTs-Setups tab.
b Click the area > cv setup.

•

This opens the capacitance vs. voltage
setup.

17 Set the LCR meter

instrument options.

a Select the Instrument Options tab.
b Highlight the Cable Length variable and

enter

2

.

•

This sets the cable length in the LCR
meter instrument options.

18 Calibrate the LCR meter.

a On the 16048D test leads, connect one tee

between the H

pot

and H

cur

connectors.

b Connect the other tee between the L

pot

and

L

cur

connectors. Select the

Measure/Simulate tab.

c Select Calibrate.
d Follow the IC-CAP prompts.

•

This performs a calibration on the LCR
meter.

•

The BNC tee part number is 1250-2405.

19 Measure the open circuit

capacitance.

a Select Measure.

•

This measures the open circuit
capacitance.

20 Observe the results

a After the calibration is complete, observe

the plot.

b Rescale the measured result (displayed in

red) cap.m.

•

The measured result should be less
than ±10E-15 farads.

* To familiarize yourself with the IC-CAP software, refer to the first three chapters of the Agilent IC-CAP 2004 User’s Guide, model number

85190D.

† To save this hardware configuration: on the IC-CAP main menu bar, choose File > Save As and enter a filename, for example

config1.hwd

(the file suffix must be

.hwd

).

‡ To save this hardware configuration: on the IC-CAP main menu bar, choose File > Save As and enter a filename, for example

config1.hwd

(the file suffix must be

.hwd

).

**S12 is identical to S21 because the attenuator is assumed to be symmetrical.

To perform the system functional verification test (continued)