Agilent Technologies 85225F User Manual

Page 77

Advertising
background image

Installing the System

2

Installation and User’s Guide

77

4 Be aware of potential

shock hazards during
floating-ground
measurements.

a Do not touch any of the DC subsystem

output connectors when the shorting bar is
disconnected and a floating-ground
measurement is in progress.

b Warn others working in the system’s

vicinity of the potential shock hazards.

•

For additional information, see the
following section

“Precautions for

Performing Floating-Ground
Measurements

.”

5 Before touching the

FORCE, SENSE, or GUARD
connectors, ensure your
safety.

a Switch off the DC subsystem and discharge

the capacitors.

b If you do not switch off the instruments,

complete ALL of the following
precautionary steps:

•

Terminate the DC subsystem
measurement by pressing the Stop key
and confirming that the MEASUREMENT
indicator is not lit.

•

Deactivate the standby mode (if used) by
pressing the Standby key and confirming
that the Standby indicator is not lit.

•

Confirm that the HIGH VOLTAGE
indicator is not lit.

•

Open the interlock connection.

6 Never use replacement

fuses with incorrect
ratings.

a After finding the cause of failure, replace

component fuses with fuses of the same
current rating and of the type specified in
the instrument’s product documentation.

•

Failure to use correctly rated fuses
could result in a fire hazard and
damage to the equipment.

7 Install the instrument so

that the ON/OFF switch is
readily identifiable and
easily reached by the
operator.

•

The ON/OFF switch is the system
disconnecting device. It disconnects
the mains circuit from the mains supply
before other parts of the instrument.

•

Alternately, an externally installed
switch or circuit breaker (readily
identifiable and easily reached by the
operator) may be used as a
disconnection device.

To ensure your safety while using the system (continued)

Step

Action

Notes