Installation and User’s Guide

A

Enhancing Measurement Accuracy

To enhance measurement accuracy

Follow these instructions to enhance the accuracy of your
measurements.

To enhance measurement accuracy

Step

Action

Notes

1 Inspect device

connections for foreign
materials or connector
damage.

a Prior to connecting the DUT, inspect the

bias network and semi-rigid RF cable
connectors for foreign materials or
damage.

b Replace any damaged connectors.

•

Before using connectors for a critical
measurement, inspect the connector
interfaces with a microscope (~20X).
Look for dirt, contaminants, dented or
scratched outer conductor mating
surfaces, and damaged center
conductors.

2 Clean the connectors.

a Remove the systems power cord from the

power supply or place the supply breaker in
the tripped position.

b Move the connectors to a well ventilated

area.

c Use a swab dipped in clean isopropyl

alcohol to cleanse the outer conductor
mating surfaces and the ends of the center
conductors.

d Blow off the remaining alcohol with clean

compressed air.

e Allow adequate time for the alcohol fumes

to disperse before activating the system.

•

Be careful not to let the alcohol get on
the insulator bead, as this may damage
the bead.

•

Be careful not to exert too much force
on the center conductors, as they may
be damaged.

•

Compressed air can reduce the
temperature of connectors
dramatically, and this can have a
significant effect upon the performance
of calibration and verification
components.

•

If the connector components being
cleaned are to be used in a critical
measurement application, allow the
temperature of these components to
stabilize prior to use.

3 Ensure proper system

warm-up time.

a Switch on line power to all of the system

components.

b Allow at least 1 hour warm-up prior to

making measurements.

•

Follow the procedure in

“To switch on

power to the system"

on page 87.

•

Do not switch off line power to the
system unless the system will not be
used for an extended period of time.

4 Ensure a proper and

constant temperature in
the measurement
environment.

a Operate the system within an ambient

temperature range of 25

°

C, ±5

°

C.

b After system measurement calibration,

hold the ambient temperature of the
measurement environment to ±1

°

C of the

ambient temperature at the time of
calibration.

•

Install heating and cooling systems as
necessary to maintain proper ambient
temperature in the measurement
environment.