Enhancing Measurement Accuracy

A

Installation and User’s Guide

113

5 Always perform a

measurement calibration
on the network analyzer
prior to making device
measurements.

a Prior to making device measurements,

complete the steps listed in

“Performing a

Coaxial System Measurement
Calibration"

on page 115.

b Ensure that the calibration kit elements are

seated correctly and firmly in the test set
cable connectors.

•

See

“Understanding System

Measurement Calibration"

on page 114

for more information.

•

Use the Agilent 85056A 2.4 mm
precision calibration kit.

•

Use the Agilent 85056D 2.4 mm
economy calibration kit.

•

Use the Agilent 85033E 3.5 mm
economy calibration kit.

•

Use the Agilent 85052D 3.5 mm
economy calibration kit.

•

Torque measurement connections
using the torque wrench provided in
the calibration kit.

6 Calibrate the system

components at regular
intervals.

a Every 6 to 12 months:

•

Perform the required Agilent E8364B
PNA Series vector network analyzer
performance verification tests.

•

Perform the required Agilent 4156C
precision semiconductor parameter
analyzer performance verification tests.

•

Perform the required Agilent 4248A
precision LCR meter performance
verification tests.

•

Perform the required Agilent E5250A low
leakage switch mainframe performance
verification tests.

To enhance measurement accuracy (continued)

Step

Action

Notes