Required tools – Agilent Technologies 85225F User Manual
Page 120

120
Installation and User’s Guide
B
DC Subsystem Functional Verification Test
Understanding the DC Subsystem Functional Verification Test
Use this procedure to manually confirm the functionality of the DC
subsystem. The procedure provided in
” confirms:
For Agilent 4156C subsystems:
• the internal operation of the Agilent 4156C precision semiconductor
parameter analyzer
For Agilent E5260A/70B DC subsystems:
• the operation of the Agilent E5260A/70B parametric measurement
mainframe
• the operation of the plug- in source monitor units
This procedure runs a self- test initiated from the instrument’s front panel.
The test includes a self- calibration routine to improve short- term
accuracy.
Required Tools
• Agilent 4156C precision semiconductor parameter analyzer,
• Agilent E5260A/70B parametric measurement mainframe