1 draining condensate, 2 venting gas, 7 setting the range using the range-setting switch – Yokogawa EJX115A User Manual

Page 36: Draining condensate -6, Venting gas -6, Warning

Advertising
background image

<7. Operation>

7-6

IM 01C25K01-01E

7.6 Venting or Draining

Transmitter Pressure-

detector Section

Since this transmitter is designed to be self-

draining and self-venting with vertical impulse

piping connections, neither draining nor venting

will be required if the impulse piping is configured

appropriately for self-draining or self-venting

operation.
If condensate (or gas) collects in the transmitter

pressure-detector section, the measured pressure

may be in error. If it is not possible to configure the

piping for self-draining (or self-venting) operation,

you will need to loosen the drain (vent) screw on the

transmitter to completely drain (vent) any stagnated

liquid (gas).
However, since draining condensate or bleeding off

gas gives the pressure measurement disturbance,

this should not be done when the loop is in

operation.

WARNING

Since the accumulated liquid (or gas) may be

toxic or otherwise harmful, take appropriate care

to avoid contact with the body, or inhalation of

vapors.

7.6.1 Draining Condensate

1) Gradually open the drain screw or drain plug

and drain the transmitter pressure-detector

section. (See Figure 7.6.1.)

2) When all accumulated liquid is completely

removed, close the drain screw or drain plug.

3) Tighten the drain screw to a torque of 10 N·m

{1kgf·m}, and the drain plug to a torque of 34 to

39 N·m {3.5 to 4 kgf·m}.

F0708.ai

Drain screw

Drain plug

When you loosen the drain screw or drain

plug, the accumulated loquid will be

expelled in the direction on th earrow.

Figure 7.6.1 Draining the Transmitter

7.6.2 Venting Gas

1) Gradually open the vent screw to vent gas from

the transmitter pressure-detector section. (See

Figure 7.6.2.)

2) When the transmitter is completely vented,

close the vent screw.

3) Tighten the vent screw to a torque of 10 N·m

{1kgf·m}.

F0709.ai

Vent screw

Vent screw

When you loosen the vent screw, the gas

escapes in the direction of the arrow.

Figure 7.6.2 Venting the Transmitter

7.7 Setting the Range Using the

Range-setting Switch

WARNING

The range setting switch must not be used in the

hazardous area.

When it is necessary to use the switch, operate it

in a non-hazardous location.

When pressure is applied to the transmitter, the low-

and high-limit values for the measurement range

(LRV and URV) can be changed (re-ranged) using

the range-setting switch (push-button) located on

the optional integral indicator plate and the external

zero adjustment screw. This procedure does

not require use of the communicator. However,

changes in the scale range and engineering unit

display settings for the integral indicator require use

of the communicator.
Follow the procedure below to change the LRV and

URV settings.

Advertising