B3. mounting, B3.1 mounting direction, B3.1 – Yokogawa YFGW610 User Manual

Page 24: B3-1, Important

Advertising
background image

<B3. Mounting>

B3-1

IM 01W02D02-01EN

B3. Mounting

Mount YFGW610 on the DIN rails being secured to the general-purpose instrumentation board or

others. No other type of mounting is allowed.

B3.1 Mounting Direction

YFGW610 is cooled down by natural air flow. Install it correctly.

Air flow

Up

Down

Air flow

FB0301

Figure B3-1 Mounting direction

IMPORTANT

• Before installing or removing YFGW610, be sure to turn its power supply off.
• Install the device to allow smooth air flow through the ventilation holes.
• To allow smooth air flow for natural cooling, secure at least 150 mm between the device’s

top panel and other devices. Be also sure to leave at least 100 mm between the device’s

bottom and other devices.

• To allow smooth routing of power supply and network cables and to avoid heat from adja-

cent devices, leave at least 50 mm between the device’s each side and other devices.

• Keep YFGW610 away from direct sunlight.

Advertising