Yokogawa FieldMate User Manual

Page 68

Advertising
background image

<6 OPERATION>

6-36

IM 01R01A11-01EN

1) For details of the calibration procedures, follow User’s Manual FY2-XJAM012-40-2E.

Record the following original parameters and set the AXF parameters as shown below

manually. (These settings are described in FY2-XJAM012-40-2E.)

Table 6.3.10 AXF parameter setting of Converter Check

HART Parameter (DD)

Parameter of

AXF Display

Setting Value

Select Flow

Tube (*1)

Device setup → Detailed Setup → Basic Setup →Se-

lect Flow Tube

C30

Calibrator

Nominal Size

Device setup → Detailed Setup → Basic Setup

→Nominal Size

C32

Enter flowtube

nominal size in

“mm” unit

Nominal Size

Device setup → Detailed Setup → Basic Setup

→Nominal Size unit

C31

mm

MF Set

Device setup → Detailed Setup → Basic Setup→MF

Set

C20

(Measure Mode)

Standard DF

or

Enhanced DF (*3)

Low MF

Device setup → Detailed Setup → Basic Setup→MF

Set → Standard DF → Low MF

C21

1.0

High MF

Device setup → Detailed Setup → Basic Setup→MF

Set → Standard DF → High MF

C22

1.0

Low MF (EDF)

(*2)

Device setup → Detailed Setup → Basic Setup→MF

Set → Enhanced DF → Low MF (EDF)

C23

1.0

High MF (EDF)

(*2)

Device setup → Detailed Setup → Basic Setup→MF

Set → Enhanced DF → High MF (EDF)

C24

1.0

*1: Only for AXFA11

*2: Only for optional code HF1 or HF2 (Enhanced Dual Frequency Excitation)

*3: Only for optional code HF1 or HF2, select “Enhanced DF”.

2) Enter the span on the “Converter” screen.

2)-1 When Verification Tool is in “online” mode:

Click “Read Span (3)” in

Figure 6.3.15. The value of parameter “Velocity Check” is entered

automatically. If the AXF parameters are different from those in

Table 6.3.10, the following

error message is displayed. Set the AXF parameters to those in

Table 6.3.10.

F060317.ai

Figure 6.3.17 Example of “Parameter Error” Dialog

Advertising