2)-2 When Verification Tool is in “offline” mode:

Read the value of parameter “Velocity Check” manually. Refer to

Table 6.3.11. And enter

the value in “cell (2)” in

Figure 6.3.15 manually.

Table 6.3.11 AXF Parameter Settings of Converter Check

HART Parameter (DD)

Parameter of

AXF Display

Velocity Check

Device setup → Detailed Setup → Basic Setup →Velocity Check

C44

3) Measure the output accuracy at the preferred setting (0%, 25%, 50%, 75%, or 100%). After

setting the value, let the AM012 stabilize at least for 1 minute.

For operation procedure, follow User’s Manual FY2-XJAM012-40-2E.

4) Enter the measured value.

4)-1 When Verification Tool is in “online” mode:

Click “Read (5)” in

Figure 6.3.15. The value of instantaneous flow rate is entered automati-

cally.

4)-2 When Verification Tool is in “offline” mode:

Read the value of parameter “PV” or “PV %rnge” manually. Refer to

Table 6.3.12, and

convert the value in “m/s” unit.

Enter the calculated value in “cell (4)” in

Figure 6.3.15 manually.

Table 6.3.12 AXF Measured Value of “Converter” Check

HART Parameter (DD)

Parameter of

AXF Display

PV

■ Process Variables → PV

or

■ Device Status → Process Variable → PV

A20

PV %rnge

■ Process Variables → PV %rnge

or

■ Device Status → Process Variable → PV %rnge

A10

5) “Error (%)” is calculated automatically.
6) “Result” is displayed automatically after filling in all fields.

F060318.ai

Figure 6.3.18 “Converter” Result

7) Set the AXF parameters back to their original values manually.