Instruction for replacement of the heater unit, 1 removal procedures for sensor assembly, Instruction for replacement of the heater unit -1 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 10: Removal procedures for sensor assembly -1, Caution

Advertising
background image

<2. Instruction for Replacement of the Heater Unit>

2-1

IM 11M12A01-21E

2. Instruction for Replacement of the

Heater Unit

2.1 Instruction for Replacement of Model ZR22A

Heater Unit

CAUTION

• Allow adequate time for the sensor (cell) and heater assembly to cool before replacing them, or

you may get burnt.

• When removing the cell, the metal O-ring should be replaced with a new one. Never reuse a

metal O-ring. Once-used metal O-ring should be replaced with a new one.

• The grease used for rubber O-rings lubricant see

, , etc. in Figure 2.2, 2.3 makes

poisonous gas when heated to more than 250ºC. Wash your hands thoroughly after work!

Especially be careful if you smoke.

2.1.1 Removal Procedures for Sensor Assembly

(1) Remove the four bolts and associated washers from the tip of the detector probe.
(2) Remove the U-shaped pipe support together with the U-shaped pipe. Remove filter also.
(3) Pull the sensor assembly toward you while turning it clockwise. Also, remove the metal O-ring

between the sensor assembly and the probe. When replacing the sensor assembly, be careful

not to scratch or dent the tip of the probe with which the metal O-ring comes in contact (the

surface with which the sensor flange also comes in contact). Otherwise, the measurement gas

will not be sealed.

(4) Use tweezers to pull the contact out of the groove in the tip of the probe.
(5) Clean the sensor assembly, especially the metal O-ring contact surface to remove any

contaminants adhering to that part. If you can use any of the parts from among those removed,

also clean them up to remove any contaminants adhering to them.

Once the metal O-ring has been used, it cannot be reused. So, be sure to replace it.

F2.1E.ai

Dust filter

(optional)

Bolts (four)

Washers

(four)

Sensor (cell)

U-shaped pipe

Metal O-ring

Contact

Probe

1/8 turn – tighten bolts 1/8 turn

(approximately 45°) each

U-shaped pipe

support

Screw

Filter

Figure 2.1 Exploded View of Sensor Assembly

Advertising
This manual is related to the following products: