<2. Instruction for Replacement of the Heater Unit>

2-2

IM 11M12A01-21E

CAUTION

If the heater strut assembly cannot be removed because a screw has fused to its thread, one of our

service representatives can fix it.

2.1.2 Replacement of the Heater Unit

1. Removing of the Sensor (Cell)
Remove the U-shaped pipe support, the U-shaped pipe, the Filter and the Sensor (Cell) according to

Sec. 2.1.1 “Removal procedures for Sensor Assembly”.
2. Removing of the Terminal Box
(1) Open the terminal box

and remove the three terminal connections (CELL+, TC+ and TC-).

(2) To disconnect the HTR terminals, remove the terminal block screw

. The other terminal

connections remain connected; disconnect the two HTR connections. (These terminals have no

polarity.)

(3) Remove the two screws

that fasten the cover and slide it to the flange side.

(4) Remove the four bolts

and the terminal box with care so that the already disconnected wire

will not get caught in the terminal box.

3. Replacement of Heater Strut Assembly
(1) Loosen screw

until heater strut assembly’s plate can be removed. There’s no need to

remove O-ring

which prevents screw from coming out. Pull out connector .

(2) Loosen and remove the screw

with a special tool (part no. K9470BX or equivalent) and then

remove the heater strut assembly

from the detector .

(3) To reassemble the heater strut assembly, reverse the above procedure:

Insert the heater strut assembly

into the detector , while inserting the calibration pipe in

the detector

into the heater section in the heater strut assembly as well as in the bracket

hole. Coat the screw

with grease (NEVER SEEZ: G7067ZA) and tighten the screw with a

special tool (part no. K9470BX or equivalent) with a tightening torque of 12N · m ±10%. (screw

attached with heater strut assembly is shipped with coating.)

4. Assembling of Connector
(1) To install the O-rings

on the calibration gas and reference gas pipes, disassemble the

connector

in the following procedure.

Remove the screw

and then remove the plate and two caps . If the O-ring remains in

the hole, pull them out from the back.

(2) Pass the heater and thermocouple lead-wire through the connector

. Also, pass the calibration

gas and reference gas pipes through the holes of the connector

.

(3) Insert the calibration gas and reference gas pipes into O-rings

after confirming no damage to

O-rings. If the O-ring

fails, replace it with a new one.

(4) Push the two caps

into the associated opening of the connector . Insert the plate ,

aligning it with the groove of the cap

, and tighten it with the screw . If you attempt to insert

the calibration gas and reference gas pipes into the connector

without disassembling the

connector

, the O-ring may be damaged.

(5) Tighten the screw

in heater strut assembly until connector cannot move.