<2. Instruction for Replacement of the Heater Unit>

2-7

IM 11M12A01-21E

2.2.1 Replacement of the Heater Unit

1. Removing of the Sensor (Cell)
Remove the U-shaped pipe support, the U-shaped pipe, the Filter and the Sensor (Cell) according to

chapter 2.11 ”Removal procedures of the sensor assembly”.
2. Removing of the Converter Case
(1) Remove the two screws

that fasten the cover and slide it to the flange side.

(2) Remove the four bolts

and the converter . Then remove the three connectors to which lead-

wire from the heater, CELL+ signal and thermocouple are connected.

If the automatic calibration unit is attached, the wires from solenoid valves are connected to the

converter with connectors.

3. Replacement of Heater Strut Assembly
(1) Loosen screw

until heater strut assembly’s plate can be removed. There’s no need to

remove O-ring

which prevents screw from coming out. Pull out connector .

(2) Loosen and remove the screw

with a special tool (part no. K9470BX or equivalent) and then

remove the heater strut assembly

from the detector .

(3) To reassemble the heater strut assembly, reverse the above procedure:

Insert the heater strut assembly

into the detector , while inserting the calibration pipe in

the detector

into the heater section in the heater strut assembly as well as in the bracket

hole. Coat the screw

with grease (NEVER SEEZ: G7067ZA) and tighten the screw with a

special tool (part no. K9470BX or equivalent) with a tightening torque of 12N•m10%. (screw

attached with heater strut assembly

is shipped with coating.)

4. Assembling of Connector
(1) To install the O-rings

on the calibration gas and reference gas pipes, disassemble the

connector

in the following procedure.

Remove the screw

and then remove the plate and two caps . If the O-ring remains in

the hole, pull them out from the back.

(2) Pass the heater, CELL+ and thermocouple lead-wire through the connector

. Also, pass the

calibration gas and reference gas pipes through the holes of the connector

.

(3) Insert the calibration gas and reference gas pipes into O-rings

after confirming no damage of

O-rings. If the O-ring

fails, replace it with a new one.

(4) Push the two caps

into the associated opening of the connector . Insert the plate ,

aligning it with the groove of the cap

, and tighten it with the screw . At this time, fix the

terminal of the end of green wire from the four core connectors to connector

by screw . If

you attempt to insert the calibration gas and reference gas pipes into the connector

without

disassembling the connector

, the O-ring may be damaged.

(5) Tighten the screw

in heater strut assembly until connector cannot move.

5. Assembling of Converter and Wiring
Reassemble in reverse order to the above disassembly procedure.

Each connector from the heater strut assembly is identified by its size and the number of cores. Insert

them fully by taking care of their size and direction. The connectors from solenoid valves are marked

“ZERO” or “Z” and “SPAN” or “S”, so correspond each marking with converter’s.
6. Installing of the Sensor (Cell)
Install the U-shaped pipe support, the U-shaped pipe, the Filter and the Sensor (Cell) according to

section 2.1.3 ”Part assembly of the sensor assembly”. heater strut assembly

is shipped with a new

contact.