Introduction – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 2

Advertising
background image

i

IM 11M12A01-21E

<Introduction>

Introduction

The ZR22A and ZR202A Heater Assemblies for ZR22G, ZR22S, ZR202G and ZR202S are described

here.
The ZR22G, ZR22S, ZR202G and ZR202S manuals are listed below.

User’s Manual

IM No.

Model ZR22G, ZR402G Separate type Zirconia Oxygen Analyzer

IM11M12A01-02E

Model ZR22G, ZR402G Separate type Zirconia High Temperature Humidity Analyzer

IM11M12A01-03E

Model ZR202G Integrated type Zirconia Oxygen Analyzer

IM11M12A01-04E

Model ZR202G Integrated type Zirconia High Temperature Humidity Analyzer

IM11M12A01-05E

Model ZR22S, ZR402G Separate type Explosion-proof Zirconia Oxygen Analyzer

IM11M13A01-02E

Model ZR202S Integrated type Explosion-proof Zirconia Oxygen Analyzer

IM11M13A01-04E

The heater assembly is a replaceable part. After unpacking, check the following. If any problems are

found, contact our sales office (do not install if problems are found).
Our warranty does not apply if user replaces the heater.

• Check after unpacking

Do a visual inspection to check that bracket is not bent, and insulator element is not chipped or broken.

• Measure resistance of heater and thermocouple (using a tester or the like):

Yellow

Green

Red

White

1

2

3

4

< ZR22A (Separate type) heater assembly >

< ZR202A (Integrated type) heater assembly >

Insulator

Insulator

Bracket

Bracket

Check that heater resistance

is 90 Ω or less.

Check that heater resistance

is 90 Ω or less.

Check that resistance of

thermocouple 5 Ω or less.

Check that thermocouple resistance

(between terminals 3, 4) of 5 Ω or less.

Media No.IM 11M12A01-21E (CD)

3rd Edition :July 2012 (YK)

All Rights Reserved Copyright © 2001, Yokogawa Electric Corporation

Advertising