IM 11M06D03-01E-A

To start calibration using an input contact, follow these steps:

(1) Make sure that Calibration start has been selected in the Input contacts display

(2) Apply an input contact to start calibration.

To stop calibration midway, follow these steps:

(1) Press the Return key. If this key is pressed midway during calibration, the calibration will stop and the output

stabilization time will be set up.

(2) Press the Return key once again to return to the basic panel display and the analyzer will be in normal

measurement.

8.5 AUTOMATIC CALIBRATION
No execution operations are required for automatic calibration. Automatic calibration starts in accordance with a
preset start day and time. Calibration is then executed at preset intervals.

Before conducting a semi-automatic or automatic calibration, run the automatic calibration unit beforehand to obtain
a calibration flow of 600 ± 60 ml/min.