III. SPECIFICATIONS, MC1

3.1 MODEL MC1, MANUAL CALIBRATION PANEL

The MC1 is the manual calibration unit for a single oxygen probe. It provides regulation of the reference air and
calibrations gas while allowing the operator to select zero or span gas for calibration. Separate flow meters are
used to set cal gas flow rates and reference air.

Figure 3.1

Maximum pressure:

35 psig

Calibration Plate:

Stainless Steel

Connections:

¼” FNPT

Calibration Tubing:

¼” Copper (standard); ¼” Stainless Steel (optional)

Weight:

Approximately 4.4 lbs. (2 kg)

Flow rates:

Calibration gas – 0.6 LPM

Reference Air – 0.8 LPM

Exceeding the recommended flow rates will damage the detector cell.