Zero Pressure Switch, M1133AR

The pressure switch is used to alert the Operator via a contact input that the zero gas cylinder needs to be re-
placed before the cylinder is empty. The recommended connection is to the high pressure side of the zero gas
regulator. Refer to Figure 4.1.

Zero Pressure Switch, M1133AR

Adjustable Set Point Ranges:

30 to 575 psi on fall; 50 to 600 psi on rise.

Dead Band:

8 to 60 psi

Over Range:

2,500 psi

Proof Pressure:

3,000 psi

Storage Temperature:

-40ºF to 180ºF (-40 ºC to 82 ºC)

Ambient Temperature:

0 ºF to 160 ºF (-18 ºC to 71 ºC)

Set Point Repeatability:

1% of span

Switch Output:

One SPDT (either NC or NO)

Electrical Rating:

Rated to 5 amps resistive and 15 amps inductive (75% PF),

at 125 and 120 VAC, ¼” HP

Enclosure:

Aluminum with irradiate finish rated for 100 hours of salt

spray

Electrical Connection:

One SPDT output; ½” NPTF, 5ft cable

Pressure Connection:

1/8” NPTM

Mounting:

NPTM pressure connection

Weight:

12oz (340g)

Approvals:

UL 508 listed, E42272; CSA C22.2, No 14-1987: LR 9690