Yokogawa AC8 Calibration Units User Manual

Page 8

Advertising
background image

8

IM 11M06D03-01E-A

3.2. SPECIFICATIONS, AC1

3.2.1 MODEL AC1, SINGLE POINT AUTOMATIC CALIBRATION UNIT

The AC1 provides control of reference air and cal gas flow rates while maintaining a constant regulation of gases.
The AC1 includes individual span, zero and block solenoids with manual overrides for easy setup of flow rates. The
user has a choice of copper or stainless steel tubing in addition to NEMA 4 or NEMA 4X enclosures. . It also contains
a 1 amp fuse with an LED indicator to indicate a power surge.

Figure 3.2

NOTE: Cutaway shows internal flow meters and solenoids

Operating Pressure: 20 psig
Maximum Pressure: 35 psig
Flow meter Range:

0.15 to 1.0 LPM (separate flow meter for cal gas and reference air).

Tubing:

1/4” Copper (standard); 1/4” Stainless Steel (optional).

Gas Connections:

1/4” FNPT

Voltage:

110 VAC, 50/60 Hz (standard).

Flow rates:

Calibration gas – 0.6 LPM

Reference Air – 0.8 LPM

Exceeding the recommended flow rates will damage the detector cell.

NOTE