3

INSTALLATION

AND
WIRING

3-2. Wiring

3-2-1. Preparation

Refer to figure 3-4. The relay contact terminals

and power supply connections are under the

screening (shielding) plate. These should be

connected first. Connect the sensor, outputs

and HART

®

communication connections last.

To open the EXAxt 450 for wiring:

1. Loosen the four frontplate screws and swing

open the cover.

2. The upper terminal strip is now visible.

3. Remove the screen (shield) plate covering

the lower terminal strip.

4. Connect the power supply and contact

outputs. Use the three glands at the back for

these cables.

• Cables that withstand temperatures of at

least 70 °C should be used for wiring.

• Wiring work should be performed to meet

IP66 or higher requirements. Tighten four

frontplate screws to 1.5 N·m torque.

Always place the screen plate over the power

supply and contact terminals for safety reasons

and to avoid interference.

5. Put back (replace) the screen (shield) plate

over the lower terminals.

6. Connect the analog output(s), the sensor

inputs, and, if necessary, the HART

®

wiring

and input contact.

7. Use the front three glands for analog output,

sensor inputs, contact input and HART

®

wiring (see figure 3-5).

8. Swing back the cover and secure it with the

four screws.

9. Switch on the power. Commission the

instrument as required or use the default

settings.

Do not turn on power with the touchcsreen

pressed, otherwise inaccurate screen

positioning will occur. If it occurs, leave the

touchscreen unpressed, turn off power then

on again. The screen positioning will be

accurate.

DANGER

Figure 3-4. Internal view of EXAxt wiring compartment

6 X M20 glands

potentio-

merter

LCD

bracket

connector for (future) software

input

terminal

block

output

terminal

block

protective

shield

bracket

Note: SC450G-A(D)-U

The enclosure is provided with stoppers in stead of M20 cable glands for the unused holes.

These stoppers must be removed and replaced by FM approved conduit fittings in accord-

ance with good installation practice. Also see Appendix 7, Control drawing for FM approval.

WARNING

CAUTION

WARNING

This connector for software must be used only by

Yokogawa’s service personnel.