Extension cable (wf10), Ba10 wf10 exa transmitter / converter – Yokogawa EXAxt SC450 4-Wire Analyzer for Conductivity/Resistivity User Manual

Page 21

Advertising
background image

13

IM 12D08N05-01E

3

INSTALLATION

AND
WIRING

3-6-1. Sensor cable connections using junc-

tion box (BA10) and extension cable

(WF10)

Where a convenient installation is not possible

using the standard cables between sensors

and converter, a junction box and extension

cable may be used. The Yokogawa BA10

junction box and the WF10 extension cable

should be used. These items are manufactured

to a very high standard and are necessary to

ensure that the specifications of the system

can be met. The total cable length should not

exceed 60 metres (e.g. 5 m fixed cable and 55

m extension cable).

14 Overall Screen

11

12

12

13

14

14

16

15

13

14

14

16

15

17

11

17

11 Red

12 Blue

13 Core 17 Screen

Brown Co-axial Cable

15 Core 16 Screen

White Co-axial cable

17 (screen)

14 (overall screen)

12 (blue)

11 (red)

13 (core)

16 (screen)

15 (core)

Co-axial cable

(white)

Co-axial cable

(brown)

11
12
13
14
15
16
17

BA10

WF10

EXA TRANSMITTER

/ CONVERTER

11

12
13
14

15
16

17

Figure 3-10.

Connection of WF10 extension cable and BA10 junction box

Other sensor systems

To connect other sensor systems, follow the general pattern of the terminal connections as listed

below:

11 and 12: Always used for temperature compensation resistor input.

13 and 14: Normally used for the outer electrode

15 and 16: Used for inner electrode

In case a 4-electrode measuring system will be used, 14 and 16 should be used for the current

electrodes. Please ensure that shielded cabling will be used.

In below figure this is shown in a schematic way.

11 12 13 14 15 16

t

11 12 13 14 15 16

t