21
IM 12D08N05-01E
5
MENU
STRUCTURE
COMMISSIONING
Menu
Parameter Default
Range
values
min.
max.
Configure Sensor
Cell constant
0.1 cm-1
0.005 cm-1
50 cm-1
Temp. Comp.
Reference Temp.
25ºC, 77ºF
0ºC, 32.0ºF
100ºC, 211ºF
Manual Comp.
Manual Temp.
25ºC, 77ºF
-20ºC, -3.0ºF
250ºC, 482ºF
Temp. Coef
T.C.methods 1
2.10%/ºC, 1.17%/ºF
-10%/ºC, -0%/ºF
3 . 5 % / º C , 2 % / º F
Temp. Coef
T.C.methods 2
2.10%/ºC, 1.17%/ºF
0%/ºC, 0%/ºF
3.5%/ºC, 2%/ºF
Note!
“Main parameter” and/or “Measure” determines the rest of the HMI menu structure
Enter
Measurement setup
Temperature settings
Temp. Compensation
Calibration settings
Concentration
Measure Conductivity only
Configure sensor
Conduct.+Concentr
Conductivity only
Resistivity only
Concentration only