7. temperature calibration, 8. operation of hold function during calibration, 9. general comments on sc calibration – Yokogawa EXAxt SC450 4-Wire Analyzer for Conductivity/Resistivity User Manual

Page 45: 8. operation of hold function during calibration

Advertising
background image

37

IM 12D08N05-01E

6

CALIBRATION

6-7. Temperature calibration

In order to make the most accurate measure-

ments, it is important to have a precise temper-

ature measurement. This affects the display of

temperature, and the output signal when used.

More important, however, is the temperature

compensation, and calibration accuracy.

The temperature of the sensor system should

be measured independently with a high preci-

sion thermometer. The display should then be

adjusted to agree with the reading (zero offset

calibration only). For best accuracy this should

be done as near to the normal operating tem-

perature as possible.

6-8. Operation of hold function during cali-

bration

EXAxt SC450G has a HOLD function that will

suspend the operation of the control/alarm

relays and mA-outputs.

During calibration, the user may choose to

enable HOLD so that the output signals are

frozen to a “last” or “fixed” value. Some users

will choose to leave the outputs “live” to record

the calibration event. This has implications

for pharmaceutical manufacture, for example,

where an independent record of calibrations is

mandatory.

Press HOLD button on mainscreen, to remove

the HOLD.

The route for HOLD setup is

Commissioning >> Output setup>> Config-

ure Hold

6-9. General comments on SC calibration

a) SC sensors experience no drift except if

they are damaged or dirty

b) There are no good calibration solutions

(like pH buffer solutions)

c) Solution calibration of SC demands

laboratory technical skills

d) Solutions can be used to give a fair

calibration check at higher conductivity

e) Solutions can NOT be used to check

calibration at low conductivity.

f) Low conductivity solutions <10

µ

S/cm

absorb CO

2

from the air very fast

g) Low conductivity measurement must be

made only with air excluded

h) Apparatus must be scrupulously clean to

avoid contamination

i) Sensor linearity is never a problem for

lower values

j) A dirty sensor is prone to polarization

k) Polarization shows as a low side error at

higher conductivity

l) A dirty sensor will often read perfectly at

low conductivity

m) Wet calibration tests are best done

towards the top of a sensor’s range

n) If the system responds correctly to the

highest trip point, all is well

Advertising