Appendix 3, Calibration solutions for conductivity, Appendix 3, calibration solutions for conductivity – Yokogawa EXAxt SC450 4-Wire Analyzer for Conductivity/Resistivity User Manual

Page 59

Advertising
background image

51

IM 12D08N05-01E

A

APPENDICES

Appendix 3

Calibration solutions for conductivity

Note!

This section should be read in conjunction with the calibration section (Chapter 6) and the mainte-

nance section (Chapter 7). The calibration (cell constant) of a sensor does not change unless the

sensor is damaged.

It can also appear to change because of coating of the electrodes, or partial blockage. Because

these changes should be handled as described in the maintenance section, it does not make

sense to regularly recalibrate the SC450G.

A calibration check, however, is another matter. When the objective is clearly defined as a diagnos-

tic exercise a regular check can bring an extra level of security and confidence to the measure-

ment.

Sensor damage, and/or coatings can be difficult to see and the calibration check can confirm their

presence, by a deviation from the known solution conductivity. The remedial action should be to

clean the sensor, and carefully check for blockage or damage (not simply to recalibrate).

Higher conductivity solutions should be used where possible. The lower the conductivity of the test

solution, the easier it is to contaminate. Carbon dioxide from the air can be quickly absorbed to

cause an error. All containers must be suitably clean, and all materials suitably pure. Outside of a

well-equipped laboratory these conditions are hard to meet.

Also note that the check must be performed with due regard to the cell constant of the sensor,

as it limits the effective working range. The documentation provided with the sensor should be

consulted to determine its maximum working value. By using a checking solution close to the

upper range limit of the sensor, the detection of contamination is better. A slightly contaminated

sensor may read perfectly at lower conductivity, but show significant errors at higher values. This

is caused by the early onset of polarization, the factor that limits the upper range of the sensor

in any case. Polarization is seen in contaminated sensors, and it is for this reason that there is a

sophisticated polarization check built into the SC450G. Details of this diagnostic tool are found in

the troubleshooting section (Chapter 8).

EXAxt SC450G is programmed with the following table of conductivity of Potassium Chloride (KCl)

solutions at 25°C. This is used in the Automatic Cell Constant setting calibration feature. (See

chapter 6 on calibration) The table is derived from the Standards laid down in “International Rec-

ommendation No. 56 of the Organisation Internationale de Métrologie Legale”.

Table 11-2. KCl values at 25 °C

mol/l

Advertising