Appendix 4

Sensor selection for the EXAxt SC450G

Yokogawa supplies the following sensors, and their application is briefly described. All are compat-

ible with the EXAxt SC450G, and the user must be aware of the sensor configuration for 2 or 4

electrodes, in order to set the converter correctly.

SC42-SP34 (&SX42-SX34) cc = 0.01 cm

-1

2-electrode sensor

This sensor is the first choice for pure and ultra pure water measurements. From 0.055µS/cm up to

100µS/cm this is a perfect choice. The measuring range extends to at least 1000µS/cm, but there

are better choices in that range.

Almost exclusively used in flow fittings and sub-assemblies. FF40 and FS40. (The -SX sensors are

screwed or flanged direct into the process pipework or tanks.)

SC4A-S(T)-XX-002...... cc = 0.02 cm

-1

2-electrode

sensor

This sensor is a good choice for pure and ultra pure water measurements. From 0.055µS/cm up

to 50µS/cm this is a good choice. The measuring range extends to 500µS/cm, but there are better

choices in that range. The “SC4A-” sensors make up a range that use compression adapters or

specialized fittings, including retractable.

SC42-SP24 (&SX42-SX24)

cc = 0.1cm

-1

2-electrode sensor

This sensor should be used for clean and pure water measurements. From 0.2µS/cm up to 1

mS/cm the sensor is fine, with measurement up to 10mS/cm possible, but there are better choices

in that range. Mounting of these sensors is as for the ....-SP(X)34 series.

SC4A-S(T)-XX-010...... cc = 0.1 cm

-1

2-electrode

sensor

This sensor is a good choice for clean water measurements. From 0.5µS/cm up to 200µS/cm this

is a good choice. The measuring range extends to 2000µS/cm, but there are better choices in that

range. The “SC4A-” sensors make up a range that uses compression adapters or specialized fit-

tings, including retractable.

SC42-EP15(D)

cc = 1 cm

-1

2-electrode sensor

This sensor is intended as a handy, short sensor. It is easy to fit in a branch (tee) on a pipeline,

and is good for measurements where the precision is less critical than a convenient installation.

The D version is particularly suited to food applications because of it’s cleanability. Range of ap-

plication is from 10µS/cm to 10mS/cm.

SC42-EP14

cc = 1 cm

-1

2-electrode sensor

Clean cooling and process water applications are where this sensor is used. The best range of ap-

plication is 10µS/cm to 10mS/cm. It is better suited to flow through installations in FS40 and FF40

adapters and fittings, but it can also be used in dip tubes - FD40.

SC42-EP18

cc = 1 cm

-1

4-electrode sensor

Cooling and process water applications are where this sensor is used, including light fouling. The

best range of application is 100µS/cm to 100mS/cm. The four-electrode system copes with some

contamination from the process, and eliminates the tendency for polarization to occur at the higher

conductivity. It is better suited to flow through installations in FS40 and FF40 adapters and fittings,

but it can also be used in dip tubes - FD40.